Radiometry is an essential technique for measuring the light output from a UV curing system in order to maintain a precise and repeatable UV curing process. The ability to accurately measure and calibrate light intensity in UV LED spot curing is a critical aspect to controlling a UV curing process but can be challenging for UV assembly manufacturers. UV measuring systems must capture the narrow beam at the center of a detecting sensor at an exact focal height in order to determine the peak irradiance.
We offer a range of radiometric and calibration devices to allow users to efficiently and accurately measure and calibrate the peak irradiance or optical power in an assembly- manufacturing environment directly with an LX500 controller or remotely with a Light Meter System (LM2011 + LS200).
Two options for measuring and calibrating UV light output are available to meet the demands of your manufacturing process, Irradiance (W/cm2) using the OmniCure LS200 or Optical Power (W) using the OmniCure LS200P. Unmatched control and repeatability can be achieved when used with an OmniCure Beam Positioning Kit and LX500 controller.
Specifications:
Measurement Range | Irradiance: 50 mW/cm2 to 40 W/cm2 Power: 1 mW to 2000 mW |
Accuracy | +/- 10% |
Resolution | Irradiance: 1 mW/cm2 Power: 1 mW |
Calibration Cycle | One Year |
Battery Life | Two Years, Typical (intermittent use) |
Operating Temperature Range | Ambient to 40° Centigrade |
Certification | Complies with IEC, Canadian and US Standards, RoHS compliant |
Functions | Peak Hold, Power On/Off, λnm (Wavelength selection), Power/Irradiance |
Traceability | NIST and NRC Compliant |
Sensor Dimension (LxWxH) | 75 mm x 25 mm x 11.95 mm |
Meter Dimension (LxWxH) | 165 mm x 100 mm x 44 mm |