The SY-TS280S UV Irradiance Meter is a high-precision UV measurement instrument featuring a separate host and interchangeable probe design. It measures UV irradiance, accumulated UV dose, temperature, and exposure time for a wide range of UV measurement applications.
Equipped with a 4.3-inch HD touch screen, the SY-TS280S provides intuitive operation and clear measurement data visualization. Its intelligent range selection automatically switches the measurement range according to UV source intensity, enabling the instrument to measure both low- and high-intensity UV sources.
The same SY-TS280S host can be used with multiple compatible probes, allowing users to select the appropriate probe according to the required wavelength range and UV application.
Key Features
- 4.3-inch touch screen for intuitive operation and clear data display
- Measures UV irradiance and accumulated UV dose
- Wide measurement range for both low- and high-intensity UV sources
- Intelligent automatic range selection
- Automatic or manual measurement unit switching
- Displays current irradiance, UV dose, temperature, and maximum irradiance
- Calculates Maximum, Minimum, Average, and Uniformity
- Uniformity Mode for evaluating UV source uniformity across different measurement zones
- Integrated high-precision temperature sensor for real-time temperature measurement
- Built-in timer for accurate measurement duration recording
- Stores up to 30 sets of measurement data
- PC software support for reading and exporting measurement data and generating reports
- Host + Probe design, allowing one host to be used with multiple compatible UV probes
Technical Specifications
| Specification | SY-TS280S |
|---|
| Irradiance Range | 0.1 µW/cm² – 100,000 mW/cm² |
| UV Dose Range | 0.1 µJ/cm² – 999,999,999.9 mJ/cm² |
| Irradiance Units | µW/cm², mW/cm², W/cm², W/m² |
| UV Dose Units | µJ/cm², mJ/cm², J/cm², J/m² |
| Unit Switching | Automatic / Manual |
| Resolution | 0.1 |
| Irradiance Accuracy | ±(5 + 5%H), where H is the measured value |
| UV Dose Accuracy | ±5% |
| Measurement Modes | General Mode / Uniformity Mode |
| Display | 4.3-inch Touch Screen |
| Data Storage | 30 sets |
| Communication | USB / RS485 |
| Operating Environment | -20°C to 60°C; humidity <85%, non-condensing |
| Host Dimensions | 132 × 83.5 × 30.2 mm |
Probe Specifications
| Specification | Probe |
|---|
| Probe Dimensions | 38 × 14 mm |
| Photosensitive Aperture | Ø8 mm |
| Operating Temperature | -20°C to 120°C |
| Material | Aluminum |
| Cable Length | 1 m |
| Extension Cable | Customizable |
Applications
The SY-TS280S UV Irradiance Meter is suitable for UV measurement in UV curing, UV LED curing, photolithography, UV coating, phototherapy, sterilization, fluorescence analysis, water treatment, research, plant cultivation, meteorology, agricultural production, and other industrial UV applications.